Largest Activity Range of any Dose Calibrator
The Atomlab 500 and Atomlab 500Plus Dose Calibrators have the largest activity range of any dose calibrator on the market today. A single ionization chamber can measure activity ranges up to 25 curies of F-18, equivalent to 100 curies of Tc-99m.
When testing for linearity, the calibrator showed a linearity deviation of less than 1.5% at 25 curies, 1% deviation at 10 curies and 0.5% deviation below 5 curies of F-18.
The Atomlab 500 and Atomlab 500Plus Dose Calibrators can accurately measure from 0.01 microcuries up to 25 curies of F-18 or 100 curies of Tc-99m.
“One of the Atomlab 500Plus features we find most helpful is that the software flows logically and is both easy to learn and simple to operate.”