| The Dose Preparation System
is used to prepare the radio-immunotherapy drug BEXXAR. It consists
of a specially designed syringe shield, vial shield and stand.
When preparing BEXXAR, the vial shield and syringe shield remain
on the stand. The stand allows the vial shield to rotate and has
a fixed stop at the 45-degree downward angle. The vial shield is
constructed of lead shielding encased in steel. The shield’s
top is threaded to allow quick insertion and removal of vials. The
top cover incorporates a pivoting shield for septum access that
allows the use of a vent needle along with the needle from the 60
cc syringe.
The syringe shield mounts on a rotating platform that slides the
syringe into the vial and locks in position. The dose can then be
drawn with forceps to minimize hand exposure. The syringe shield
is constructed of .84" lead shielding encased in steel. A high-density
lead glass window allows viewing up to the 50 cc mark on a 60 cc
syringe. When the syringe is filled and assayed, it can be transferred
to the BEXXAR Unit Dose Pig for safe transport.
The Dose Preparation System provides a safe and shielded environment
while making it easy to prepare a BEXXAR dose.
007-978 Dose Preparation Syringe Shield
- Dimensions: 3" dia x 5.6" l (7.6 x 14.2
cm)
- Lead Shielding: .84" thick (2.1 cm)
- Lead Glass Window:
- Dimensions: .50" w x 4.25" l x .38" thick (1.27
x 10.80 x 97 cm)
- Density: 5.6
- Construction:
- Exterior: Steel
- End Cap Mechanism: Stainless Steel
- Finish: Powder coat
- Weight: 13 lb (5.9 kg)
001-850 Dose Preparation Vial Shield
- Dimensions:
- Exterior:
6.35" h x 4" dia (16.1 x
10.2 cm)
- Interior: 3.35" h x 1.17" dia (8.5 x 2.9 cm)
- Lead Shielding:
- Cylinder: 1.17" thick (3 cm)
- Bottom End Cap: .81"thick (2.1 cm)
- Top End Cap: 1" thick (2.5 cm)
- Cover: .75" thick (1.9 cm)
- Construction:
- Exterior: Steel
- Interior: Molded lead
- Top End Cap: Aluminum
- Bottom End Cap: Plastic
- Finish: Powder coat
- Weight: 21 lb (9.5 kg)
042-455 Dose Preparation Stand
- Dimensions: 12.6" w x 10.6" d x 12.6"
h (32 x 26.9 x 32 cm)
- Construction: Stainless Steel and Steel
- Finish: Stainless Steel and Powder Coat
- Weight: 21 lb (9.5 kg)
 |
 |
| 042-460
|
Dose Preparation
System
Includes: 007-978 Dose Preparation
Syringe Shield, 001-850 Dose Preparation
Vial Shield and 042-455 Dose Preparation Stand. |
$2,070.00
|
|
 |
Components:
 |
 |
| 007-978
|
Syringe Shield,
Dose Preparation,
BEXXAR, 60 cc |
$750.00 |
|
 |
| 001-850
|
Vial Shield, Dose
Preparation,
BEXXAR, 30 ml |
475.00 |
|
 |
| 042-455
|
Dose Preparation
Stand |
995.00 |
|
 |
Related:
 |
 |
| 066-536 |
Forceps, Curved,
non-locking,
13.5 l" (34.3 cm) |
$48.00 |
|
 |
|
|


Syringe Shield mounts to the stand on a stainless steel rotating
platform; syringe is placed in shield. Note Vial Shield in upright
position.

The Vial Shield rotates to 45-degree
downward angle. Swing open the cover
for Vial Septum access.

Syringe Shield slides upward allowing the syringe to puncture the
Vial Septum.
Dose is drawn.
|