| The PET-CT Phantom™ includes
internal structures (three rods and six spheres) which, when
imaged with both modalities, can demonstrate how accurately the
two image sets are aligned.
In addition, a single sample of radioactive
water is attenuated by water, bone and CT contrast material (as
well as air only) to determine how accurately the CT-based PET
attenuation correction works.
The Phantom is used for:
– Acceptance testing of PET/CT and SPECT/CT systems
– Routine quality evaluation of PET/CT and SPECT/CT systems
– Evaluation of new image fusion software
– Evaluation of new attenuation correction algorithms
– Aluminum tubes are for registration
– The outer 2" OD micro cylinder is for comparing attenuation region
to non attenuation region
– The 6" ring is for contrast solution
– Research
Specifications
- Main Cylinder:
Interior length of phantom: 180
mm
Fillable spheres (5) inner diameter: 10 mm, 13
mm,
17 mm, 22 mm, and 28 mm.
- Distance from sphere plane to inside wall: 70 mm
- Volume of empty cylinder: 9.7 liters
- Main Cylindrical insert dimension:
Outside diameter: 51 mm
Length: 180 mm
- Top Cylinder:
Cylinder outside diameter: ~5.1 cm
Cylinder inside diameter: ~4.
cm
Cylinder inside height: ~8.2 cm
Cylinder outside height: ~12.0
cm
Volume of empty cylinder: 408 cm2
- Three Aluminum Tubes:
One 5 inch long: ~1.7 cc
Two 7 inch (ea): ~ 2.5 cc
- Stepped Bone Ring:
Pre-filled with liquid bone composition, not to be opened
The
volumes for the bone ring are:
Outer volume: 15.6 cu inch: ~256 cc
Inner volume: 6.7 cu inch:
~110 cc
 |
| 043-771 |
Phantom, PET-CT |
$4,290.00 |
|
 |
|
|
 |