Atomlab Dose Calibrator Physics Tests and Reports

 

Results on the Tests Performed on the New Dose Calibrator from Biodex Medical Systems
Submitted by Elan Gandsman

Tests of Precision and Accuracy
Tests of Linearity

Conclusions:

The dose calibrator performed extremely well in all tests.

The precision and accuracy tests for a wide range of radioisotopes showed that the precision for individual measured activities was well within ±5% for activities over 1 µCi. Even for very low activities (0.38 µCi) the precision was within ±10%! The accuracy was within ±10% even for the lowest measured activities. For higher activities it approached the 1% level.

The linearity of the instrument was better than ±1% for the measured range of activities (291 mCi to 11.8 mci). No saturation effects were observed. The geometry test failed to detect a significant change in measured activity where the volume changed from 1cc to 10cc.

Especially remarkable is the fact that the dose calibrator was capable of measuring very small activities of different radioisotopes with a high degree of precision and accuracy.

Tests of Precision and Accuracy

 

Tests of Linearity

 