Home Nuclear Medicine
Atomlab Dose Calibrator Physics Tests and Reports


Dose Calibrator Performance Evaluation of Atomlab 100
Performed by Kenneth A. Coleman, M.E. Radiological Physicist

Dose Calibrator Linearity Tc-99m Decay Method
Daily Constancy Check
Dose Calibrator Accuracy Evaluation
Activity Linearity Evaluation Using Calicheck System
Geometrical Variation Evaluation
Horizontal Variation Evaluation
Comparative Reading Log
Vertical Variation Evaluation

Conclusions:

A week-long testing of the Atomlab 100 Dose Calibrator resulted in the following conclusions.

The tests included Dose Calibrator Linearity testing from 6-12 February 1989, Dose Calibrator Constancy Tests over the same dates, Activity Comparisons for many radionuclides over the same period (this data not analyzed but provided for information), Dose Calibrator Linearity Tests by the Attenuation Method (using the Calicheck) performed Monday through Thursday, Geometrical Evaluation, Vertical Variation Evaluation and Horizontal Variation Evaluation.

The Dose Calibrator Linearity evaluation showed an excellent linearity between 500+ mCi down through less than 10 µCi; the daily linearity test fell apart on the fourth day when the activity in the shield of greatest thickness fell below 1 µCi. I am not concerned by this performance since neither the NRC or any user I know would ever expect a unit to function over this range. The Calicheck Factor Calculation is provided to show the range of correction factors that would have resulted if the initial values were calculated that day. The Dose Calibrator functioned well within the requirements for linearity.

The Dose Calibrator Accuracy tests show that the unit functions well within the required standards.

The Geometrical Variation Test indicated a variation well within the applicable standards.

The test for Vertical Variation is an assessment of the change in reading as a source is moved vertically along the axis of the chamber. The system performed in the expected manner with a slight increase as the source neared the geometrical center of the chamber and a significant fall off as the source left the chamber. The system functions well.

Daily Dose Calibrator Constancy showed an acceptable response even though the system read the source consistently higher than its calibrated activity.

The test for Horizontal Variation shows the effect of moving the source across the dipper well and the system functioned well.

Regarding the unit I tested, I offer the following comments:

1) It is an excellent decision that the chamber and display unit are calibrated separately and either can be returned for repair or calibration as needed.

2) It would be helpful (and different) if the circular depression milled in the cup at the bottom of the dipper was made slightly larger in diameter (not deeper) in order to allow the sealed sources used in daily constancy tests to fit more securely in the base. This might decrease any effect of horizontal variation when the source tips over in the dipper or is not placed in the hole because it doesn't sit well anyway.

My conclusion after thoroughly testing this unit is that this system is an excellent and needed addition to the Nuclear Medicine armamentarium and I look forward to being able to provide this unit to clients of Radiation Services.

Dose Calibrator Linearity Tc-99m Decay Method

Daily Constancy Check

Dose Calibrator Accuracy Evaluation

Activity Linearity Evaluation Using Calicheck System

Geometrical Variation Evaluation

Horizontal Variation Evaluation

Comparative Readings Log

Vertical Variation Evaluation

 


Home Nuclear Medicine

 
Home About Biodex Support Services Contact Biodex Terms of Use