- All functions microprocessor controlled
- Ultra-fast response
- Automatic range selection Ranges up to 9.999 Curies of
Tc-99m or 2.5 Curies of F-18
- Pre-programmed for the most commonly used radionuclides
- Display in Curies or Becquerels
- Automatic background subtraction and zeroing
at the touch of a button
- Remote ionization chamber with .25" (.64 cm)
lead shielding and 10-foot (3.05 m) cable
- Self-diagnostic software
- Daily constancy isotope keys
- Electronic power supply (no battery in chamber)
- Industry exclusive two-year warranty
- Inventory control of 25 samples, correcting for
volume, activity, and moly concentration
- Volume determination and future dose computations
- Pharmaceutical quality control programs
- Isotope decay projection
- Automatic linearity calculations using
attenuator tubes
- RS-232 bi-directional serial communications port
- Dot matrix printer
- ETL to UL 3101-1 and cETL to CAN/CSA
C22.2 No. 1010.1-M92
The Atomlab™ 200 provides fast, accurate radionuclide
activity measurements with performance that easily surpasses the
most stringent regulatory requirements.
The unit is simple to operate. There are 13 isotope selection keys,
10 are pre-programmed for the most commonly used radionuclides and
3 are user defined. There are 94 isotope-specific dial values listed
in the manual, including Y-90 and Sr-89. Any key can be reprogrammed
by the user for a desired isotope.
Activity is displayed on a LED readout in either Curie or Becquerel
units. Background correction and zero adjustment are performed at
the touch of a button. Range selection is automatic.
Activity measurements are performed by a microprocessor controlled
electrometer located within the detector assembly of the ionization
chamber. The chamber is shielded with .25" (.64 cm) lead. It
can be located up to ten feet away from the display unit. Chamber
bias is generated within the display unit by an electronic high
voltage supply, eliminating the need for expensive battery changes.
The Atomlab™ 200 Dose Calibrator provides inventory management, quality
assurance, and record-keeping functions. The dot matrix printer
allows hard copy records to be produced for all functions. The microprocessor-driven
memory provides inventory control for 25 samples, storing and correcting
the volume, activity, and moly concentration. When drawing a dose
from an inventory sample, the Atomlab™ 200 will display the required
volume for a specific dose and, if required, print a decay chart
showing required volumes for that dose at future times.
The Atomlab™ 200 facilitates dose calibrator quality
assurance with automatic constancy and linearity functions. The
system stores and decay corrects Co-57 and Cs-137 reference sources
and compares the measured activity to the calculated activity for
the daily constancy test. Quarterly linearity tests are simplified
by software which guides the user through the attenuator tube procedure
(regardless of which brand you use), stores all values, and does
all calculations.
For radiopharmaceutical quality control, the Atomlab 200 performs
all counting and calculations for paper chromatography tests, computing
the percentages of free pertechnetate, hydrolyzed reduced Tc-99m,
and labeled radiopharmaceuticals.
The RS-232 port enables the Atomlab™ 200 Dose Calibrator to communicate
with most commercially available nuclear medicine management systems.
An industry exclusive two-year warranty is standard.
- Isotope Selection Keys: Ten pre-programmed —
Tc-99m, Mo-99, Tl-201, I-123, Xe-133, Ga-67,
In-111, I-131; Cs-137, and Co-57; 3 user-defined keys
- Activity Range: 0.01 µCi to 9999 mCi (.001 MBq to 399.9
GBq) of Tc-99m
- Energy Range: 25 keV to 3 MeV photons
- Response Time: One second for doses greater than 2 mCi; Three
seconds for doses between 200 µCi and 2 mCi; 3-30 seconds
for doses below 200 µCi
- Detector Linearity:± 1% or 0.2 µCi, whichever is
greater
- Electrometer Linearity: ± 1% or 0.2 µCi, whichever
is greater
- Electrometer Accuracy:± 1% or 0.2 µCi, whichever
is greater
- Overall Accuracy:± 3% or 0.3 µCi, whichever is greater.
Overall accuracy is affected by such factors as the accuracy of
the specific source calibration, geometric variations due to sample
volume or configuration, detector linearity, electrometer accuracy
and readout accuracy
- Repeatability:± 0.3% above 1 mCi short term (24 hr); 1%
long term (1 yr)
- Digital Calibration Dial: Four-digit LED dial display with increment/decrement
keys to change the value. Range is from 0.1 to 999
- Detector: Well-type pressurized ionization chamber, with Argon
fill gas; Well opening 2.75" (7cm), well depth 10.5"
(26.7 cm)
- Detector Shielding: .25" (6.3 mm) lead on all sides except
top well opening. Supplementary shielding available.
- Chamber Bias: 340 volt electronic power supply
- Environmental Operating Conditions:
-
Temperature: 0-40° C;
-
Humidity: 0-90%
- Power Requirements: 100 to 120 VAC @ 1/2A; 200 to 240 VAC @ 1/4A
- Line Frequency: 50/60 Hz; Detachable line cord; Built-in EMI filter
and transient suppression
- Detector and Interface Cables: 10' (304.8 cm) long, six conductor
cables (two carry power, one is chassis ground, three carry serial
data for digital I/O) Ends terminated with AMP Mate-n-Lock connectors,
white with hooded strain relief
- Display Unit:
- Dimensions: 12" w x 14.3" depth x 3.75" h (30.5
x 36.3 x 9.5 cm)
- Weight: 6 lb (2.7 kg)
- Detector Unit:
- Dimensions: 7.5" w x 7.5" depth x 16.25" h (19
x 19 x 41.3 cm)
- Well I.D.: 2.75" dia x 10.5" h (7 x 26.7 cm)
- Well I.D. with Liner: 2.5" dia x 10.25" h (6.35 x 26
cm)
- Lead Shielding: .25" lead (.63 cm)
- Weight: 35 lb (16 kg)
- Approvals: ETL to UL 3101-1 and cETL to CAN/CSA C22.2 No.1010.1
M92
 
 |
 |
| 086-270 |
Dose Calibrator,
Atomlab™ 200, 115 VAC
Includes RS-232 port, Vial/Syringe Dipper,
Well Insert and Printer |
$7,000.00 |
|
 |
| 086-297 |
Dose Calibrator,
Atomlab™ 200, 230 VAC
RS-232 port
Includes RS-232 port, Vial/Syringe Dipper,
Well Insert and Printer |
7,000.00 |
|
|
Optional:
 |
 |
| 086-301 |
Shielding Rings,
Interlocking, 2" lead For
additional protection from
high energy activity |
$2,300.00 |
|
 |
| 086-423 |
Moly Shield, Vial,
.3" lead |
125.00 |
|
 |
| 086-435 |
Moly Shield, Syringe,
.3" lead |
425.00 |
|
 |
| |
Replacement:
 |
 |
| 086-242 |
Vial/Syringe Dipper |
$70.00 |
|
 |
| 086-241 |
Well Insert |
70.00 |
|
 |
| 086-278 |
Ribbon, Printer,
4/pkg |
18.00 |
|
 |
| 086-277 |
Paper, Printer,
1 part, 6 rolls/pkg
165' per roll. |
25.00 |
|
 |
| 086-279 |
Paper, printer,
2 part, 6 rolls/pkg
100' per roll. |
40.00 |
|
|
|