| For facility entrances,
waste management areas, inter-department corridors
- Detects gamma radiation
- 0.1 mR/hr - 1 R/hr range
- Digital display
- Includes two wall-mount NaI (Tl) scintillation detectors
with enclosures and cables
- Separate LOW and HIGH alert/alarm with yellow/red
lights, adjustable audible tones
- Detector failure light and tone to signal
detector overload or instrument failure
- Low battery warning
- Threshold adjusts from 2 to 100 mV
- Battery continuously trickle charges
when unit is operating and connected to wall current
This Area Monitor serves as a radiation sentry/alert for
building entrances, halls, and doorways. Equipped with a pair of NaI
(Tl) scintillation detectors that mount on both sides of a doorway or
hall, the unit continuously monitors incoming and outgoing materials
and people for low-level gamma radiation. Typical applications are the
monitoring of areas where waste material leaves departments or buildings,
and monitoring emergency room entrances in the event of radiation-related
emergencies. The Area Monitor can be set to give a visual and/or audible
alert (yellow light, slow beep) when a selected background level has
been exceeded, and give an alarm signal (red light, fast beep) when a
selected radiation level is reached. Lights and audibles also indicate
detector overload, no count from detectors, or instrument failure. A
yellow LED illuminates when the battery has less than 2 hours of power
remaining. The monitor is wall-mountable. Two 25-foot detector connection
cables are included.
- Detectors (2): 3" dia x 1" thick
(7.6 x 2.5 cm) shielded NAI (Tl) scintillation detectors in
NEMA4x enclosures
- Display: Four-digit
LED display with 0.8" (2 cm) character
height
- Display Units: Can be made to display in µR/hr, mR/hr,
R/hr, µSv/h,
mSv/h, Sv/h, µrem/hr, mrem/hr, rem/hr, cpm, cps, and others
- Linearity:
Reading within ±10% of true value with detector
connected
- Response: Typically three seconds from 10% - 90% of final
reading
- Status: (green light) instrument functioning properly
- Alarm: Low
Alarm - indicated by yellow light and slow beep (one per second)
audible tone (can be set at any point from 0.0 - 9999); High
Alarm - indicated by red light and fast beep (four per second)
audible tone (can be set at any point from 0.0 - 9999); NOTE:
Audible indicators can be configured as a single beep if desired
- Det Fail: Detector
overload, no count from detector, or instrument failure Ð indicated
by red light and audible tone (greater than 68 dB at 2 feet)
- Low
Bat: (yellow) indicates less than two hours of battery power
remaining
- High
Voltage: Adjustable from 200 - 2500 volts
- Threshold: Adjustable
from 2 - 100 mV
- Dead Time: Adjustable to compensate for dead time
of detector and electronics (can be read on display)
- Overload: Senses
detector saturation (indicated by display reading :"-OL-")
- Overrange:
Indicates radiation field being measured has exceeded counting
range of instrument (indicated by display reading "-
- - -")
- Data Output: Nine-pin connector providing five decade
logarithmic output, RS-232 output, signal ground connection,
FAIL and ALARM signals (current sink), and direct connection
to battery and ground
- Calibration Controls: Accessible from front of instrument
(protective cover provided)
- Power: 115 VAC; 220 VAC or battery power
- Battery Life: Typically
48 hours in non-alarm condition, 12 hours in alarm condition
- Battery
Charger: Battery is continuously trickle charged when instrument
is connected to line power and turned on
- Battery Dependence: <3%
change in readings to battery endpoint
- Monitor Construction: Wall
mount aluminum housing with ivory polyurethane enamel paint
- Dimensions:
- Monitor electronics unit: 7.4" h x 9.7" w x 2.5" depth (18.7
x 24.6 x 6.4 cm)
- Detectors (each): 13" h x 17" w x 8.5" l (33
x 43 x 22 cm)
- Connecting cables (each): 25-foot, type "C" connector
- Weight:
- Monitor electronics unit: 6.5 lb (2.3 kg)
- Detectors (each): 32 lb (14.5 kg)
 |
| 051-022 |
Monitor, Area, Entrance/Exit,
115 VAC
Includes: Two NaI (T1) Scintillation Detectors |
$5,895.00 |
|
 |
| 051-024 |
Monitor, Area, Entrance/Exit,
230 VAC
Includes: Two NaI (T1) Scintillation Detectors |
5,895.00 |
|
|
|

 |